DoseAware Radiation Dose Tracker

DoseAware

Radiation Dose Tracker

Product number:

722360

Building a strong radiation safety culture starts with giving the medical staff real-time insight about their personal exposure. DoseAware is an active dosimetry system that provides real-time insights about radiation exposure, helping medical staff and physicians evaluate and directly adjust their behaviors.

Specifications Show all

Personal Dose Meter
Weight
30 grams (1.06 ounces)
Operational unit
30 grams (1.06 ounces)
X-ray dose range
1 μSv – 10 Sv
X-ray dose resolution
1 μSv
X-ray dose uncertainty
5% or 1 μSv
X-ray dose rate range and linearity
+/- 10% 40 μSv/h – 150 mSv/h
+/- 20% 150 mSv/h – 300 mSv/h
Energy dependency rays
N40-N120 (33 keV – 101 keV)
Average battery life
3 – 5 years, depending on daily use
DoseAware Xtend Hub (to be mounted on ceiling)
Dimensions
13.5 x 18.5 x 3.5 cm (w x h x d) / 5.3 x 7.3 x 1.4 inch
Weight
170 grams (6 ounces)
Display
via FlexVision XL
Resolution
640 x 480 pixels
Storage
All X-ray dose-rate/sec and accumulated X-ray dose/hr that are received from PDMs in range.
For example: the memory size accommodates 250 PDMs with 50 hours X-ray dose rate history each.
Communication
Wireless radio communication with PDMs (868,3 918,3 927,9 MHz)
Ethernet
10/100 Mbits/s port for the Dose Manager and connection to X-ray system.
DoseAware Xtend reporting
Email reports
By week or by month, by PDM or by lab
Operator RDSR
Comprehensive Structured Report, up to 3 SCP can be configured as receivers.
Operator RDSR SOP class
UID 1.2.840.10008.5.1.4.1.1.88.33
DoseAware Base Station
Dimensions
30 x 25 x 6 cm (w x h x d) / 11.8 x 9.8 x 2.4 inch
Weight
1.45 kilograms (51.15 ounces)
Display
10.4 inch touch screen
Resolution
640 x 480 pixels
Storage
All X-ray dose-rate/sec and accumulated X-ray dose/hr that are received from PDMs in range.
For example: the memory size accommodates 250 PDMs with 50 hours X-ray dose rate history each.
Communication
Wireless radio communication with PDMs (868,3 918,3 927,9 MHz)
Ethernet
10/100 Mbits/s port for the Dose Manager connection
Dose Manager (PC requirements)
Operation system
Windows XP, Vista or Windows 7
System memory
At least 2 GB
Hard disk
40 GB with at least 15 GB available space - USB 2.0 port
DoseView (PC requirements)
Operation System
Windows XP, Vista or Windows 7
System memory
At least 1 GB - USB 2.0 port